Shenzhen Bestman Instrument Co., Ltd.

home products Others Multi-parameter Patient monitor

Multi-parameter Patient monitor

Product Details:

NIBP monitor patient monitor
Place of Origin Guangdong, China (Mainland)
Model Number BPM-9000ABC

Payment & Shipping Terms:

Price: FOB Shenzhen USD 1301

Detailed Product Description


Multi-parameter patient monitor is used for inspecting the patients on body features such as ECG,NIBP, HR,SpO2,RESP,TEMP;

Measuring synchronously 3-7 lead of ECG, HR, SpO2, RESP, TEMP, PR, NIBP and pulse conducting time etc;

12.1'' TFT Big-screen, real-color, wide-visual-angle and high-brightness LCD display;

Built-in rechargeable batteries with maintenance-free and high-capacity can work more than 2 hours without additional power supply;

With the linked formed by network, bed-side machines and central multi-parameter monitor system, constituting the web of monitoring;

Able to display trend data, and manual printing and alarm trigger printing functions available;

Able to use together with high-frequency electrical cutter and defibrillator.

Specifications:
I .Safety level
Class I anti-electroshock equipment, ECGRESP, TEMP--Type CF, SPO2, NIBP--BF

2. Power supply
AC 100-250V , 50/60Hz, Pmax=60VA, FUSE 0.5A

3. Display
True color,12.1''  type LCD, resolution 640x480

4. Signal interface
External display interface: standard VGA

External communication interface: RS-485

External print interface: standard combining-port

5. Recorder
Recording width: 48mm

Paper speed: 25mm/s

Trace: 3 channel

6. ECG specifications:
Lead mode:12 leads R, L, F, N, C1, C2, C3, C4, C5,C6

Lead selection: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6

Waveform:12 channels

Gain: ×0.5, ×1.0, ×2.0

Modes: diagnosis, monitoring, operation

Input impedance: >5MΩ

CMRR: >90dB

Electrode offset potential: ±300mV

Leakage current :<30μA

ECG signal range: ±10Mv (Vp-p)

Frequency response (bandwidth): Diagnosis 0.05-100Hz

Monitoring: 0.5-40Hz

Operation: 1-20Hz

Measurement range: 10-254bpm

Resolution: 1bpm

7. RESP specifications:
Method: Impedance between RA-LA

Input impedance: >2.5MΩ

Bandwidth:0.3-2Hz

Resp.rate:0-120rpm

Measuring Range: 0-120rpm

Resolution: 1bpm

8. NIBP specifications:
Method: oscillation

Mode: manual, periodic, continuous

Measuring interval in periodic mode: 3, 10, 45, 60 min

Continuous mode:5 min

Pulse rate range:40-240bpm

Overpressure protection: adult 290 mmHg child 145 mmHg

9. SPO2 specifications:
Measuring range:0-100%

Resolution: 1%

Actualization interval: 1sec.

Alarm delay: 10 sec.

Pulse rate: measuring range 0-254bpm

10. TEMP Specifications:
Measuring range: 0-50

Resolution: 0.1

Actualization interval: 1sec.

Average time constant :<10 sec.

Multi-parameter Patient monitor Multi-parameter Patient monitor

Send your message to this supplier
« Prev Product 1 2 3 4 5 6 7 ... Next Product »